EClerk-M-2Pt-G3-HP High Accuracy Temperature Transport Data Logger

• Stores a Total of 250,000 Recorded Temperature Readings for Each Channel 
• Violation Fixation Function

• User-Configurable Alarms
• LED Indicator

• -50…+200°С Temperature Measurement Range
• Temperature Reading Accuracy±(0,2+0,001Т)°С
• Interface to connect USB-cable for Set-up and Data Download
• Terminals for two-wire connection of two sensors Pt1000
• Installation in the driver's cabin for real-time temperature monitoring
• Mobile Application EClerk2.0-mobile  
 
 

Request

Share on Social Networks
  • Good solution for long-term measurements with high accuracy for monitoring cold chain transportation
  • Terminals for two-wire connection of two sensors Pt1000
  • Installation in the driver's cabin for real-time temperature monitoring
  • With or without LED display
  • With or without violation fixation function
  • Data logger with violation fixation function has protection against data changes
  • Downloading data from the Transport Data Logger to the phone directly at the place and sending them by e-mail
  • Print report in Excel formatInstallation in the driver's cabin for real-time temperature monitoring

Installation in the driver's cabin for real-time temperature monitoring

Downloading data from the Transport Data Logger to the phone directly at the place and sending them by e-mail

Choose your modification

Part Number Model Violation Fixation LED Indicator
1 052330001 EClerk-M-01-2Pt-G3-HP --- ---
2 052330011 EClerk-M-11-2Pt-G3-HP ---
3 052330101 EClerk-M-01-2Pt-G3-HP-a ---
4 052330111 EClerk-M-11-2Pt-G3-HP-a

Memory capacity (number of recordable values) 250000 values for  each channel
Number of channels 2
Measurement range -50…+200°С
Accuracy ±(0,2+0,001Т)°С
Temperature reading resolution 0,1°C for display, 0,03°C for soft
Real time clock ±1 sec per day 
Sampling interval 1 sec … 24 h
Data record type cyclic, till filling up
Start type by time, by button
Additional function mode of operation "daily cycles"
Number of recording intervals (sessions) up to 21
Power supply from an 1/2AA element with a voltage of 3.6V or from USB
Protection class IP54
Size 145 х 90 х 41 mm
Operating temperature –40…+55°С
Case material ABS
Recording capacity for sampling interval
  • 70 hours for 1 sec
  • 30 days for 10 sec
  • 180 days for 1 min
  • 30 year for 1 hour
Battery life for sampling interval
  • 210 days for 1 sec
  • 1,9 years for 10 sec
  • 2,6 years for 1 min
  • 2,7 years for 1 hour


Accessories*

1/2AA battery for EClerk-M data logger

*Accessories is specified in the field for the order together with the selected data logger or independently (if only additional equipment is ordered).





Software

EClerk-2.0

Free Soft for Windows

Configuring
Data analysis in table and graph
Filters by time, filters by maximum and minimum value
Report
Export data to Excel
Online mode (USB meter)
Download
EClerk-2.0 mobile

Free Soft for Android

Data download from data loggers to Android device and then transfer to remote PC via GSM
Download on Google Play